c.
Leave the banana style connector loose, ensuring that it is not contacting any
other connectors.
3. Connect the HT sample stage temperature control and sensing cable (labeled
DC2048) to the rear panel of the Model 336 temperature controller as follows:
a. Connect the DIN style connector (labeled Sensor) to Input D.
b. Connect the banana style connector (labeled Heater) to Output 1 (heater).
c.
Connect the green ground banana plug from the banana plug to the screw
terminal adapter.
4. Locate the other end of the HT sample stage temperature control and sensing
cable (labeled DC2048). It is a 6-pin connector. Attach it to the rear of the vacuum
chamber, which has a corresponding 6-pin connection. Note the orientation of
the guiding lugs, and rotate as necessary. The outer shell will rotate and click into
place for a secure connection.
5.2.2.1.2 Reconfigure the Model 336 Temperature Controller for HT Sample Stage Operation
1. Verify that Input D of the Model 336 temperature controller is properly config-
ured for the HT stage as listed in TABLE 5-1.
2. Reconfigure Output 1 according to TABLE 5-2.
3. Conduct the temperature sensor and heater test as outlined in section 3.7.2 on
Output 1 and Input D to verify that all connections and configurations have been
made properly.
If the controller is configured improperly, the controller can provide enough heat to
damage the probe station.
FIGURE 5-1
Connect the cables to the
Model 336 temperature controller
FIGURE 5-2
Connect the other end of the HT sample stage cable