63
|
www.lakeshore.com
3.6.4 Assembling the
Vision System
FIGURE 3-17 illustrates the assembly of a single-post microscope, CCD camera and
ring light. The vertical post, horizontal boom, microscope and CCD camera are
connected to the Model CRX-6.5K baseplate. The position of the shaft collar
determines the height of the microscope above the sample, and may require
adjustment after assembly of the vision system.
3.6.4.1 Assembling the Vertical Post
Follow this procedure to install the vertical post onto the baseplate. Depending on
the microscope option and probe station model, the microscope post will have either
one or two shafts. The Zoom 70 has one shaft and the Zoom 160 has two.
1. Install the microscope vertical post as shown in FIGURE 3-18. To install the post,
remove the one or two M5 screws in the bottom of the post.
2. Align the post so that one of the two screw holes on the vertical post mount is
over the hole that is closest to the vacuum chamber. The other screw hole will
align with the third hole further away from the vacuum chamber.
3. Place the two M5 screws in their respective holes and finger tighten each one.
4. Using the 5 mm hex driver, tighten each one until secure.
FIGURE 3-17
Assembly of the vision system onto the probe station
Hand dial
Microscope
Ring light
Viewport
Chamber
Monitor
Horizontal boom
(swing arm)
S-video
cable
Color CCD
camera
Optical
fiber cable
Light
source
Vertical
post