64
CHAPTER 3:
Installation and Setup
Model CRX-6.5K Probe Station
3.6.4.2 Assembling the Microscope and Horizontal Boom
1. Using the 3 mm hex driver, remove the four M3 mounting screws (FIGURE 3-18)
from the microscope.
2. Attach the microscope to the horizontal boom (see FIGURE 3-18) with the screws
that were removed in step 1. The microscope attaches beside the hand dial.
3. Slide the horizontal boom onto the vertical post as shown in FIGURE 3-19. You
may need to turn the quick release handle counterclockwise to allow it to slide
easily. If the microscope physically touches the vacuum chamber lid, adjust the
shaft collar before moving on; section 6.3.4.5 describes how to adjust the shaft
collar.
FIGURE 3-18
Left: Installing the microscope vertical post; Middle: Four M3 mounting screws on the vertical post;
Right: Attaching the microscope to the horizontal boom
FIGURE 3-19
Installing the horizontal boom onto the vertical post