74
CHAPTER 3:
Installation and Setup
Model CRX-6.5K Probe Station
You only need to loosen the screws in steps 3 to 5. Do not remove them. This will keep
them from dropping into the chamber.
3. Using the 1.5 mm hex driver, loosen both of the M3 probe arm set screws two to
three turns (FIGURE 3-34) to rotate the probe to access one of the probe mount
set screws.
4. Using the 1.5 mm hex driver, loosen the visible probe mount set screw
5. Rotate the probe mount to access the set screw on the other side of the probe
mount and loosen that screw as well.
6. Gently work the blade up and down to release and remove it. Avoid sideways
force to minimize the chance of cracking the alumina.
7. Remove the blade from the system and place it back into its storage case, making
sure not to contact the delicate probe tip.
FIGURE 3-33
Disconnect the SMA
FIGURE 3-34
Loosen the probe arm set screws
FIGURE 3-35
Loosening the probe mount set screws