100
CHAPTER 4:
Basic Operation
Model CRX-6.5K Probe Station
Probe station cool down
to base temperature
Maintaining sample
stage at base temp
Operating sample
stage 10 K t0 350 K
Returning probe station
to room temperature
CCR control
settings
Main power
On
On
On
On
On/Off
On
On
On
Off
Electronic
control
settings
Sample stage
Heater range
Off
Off
High (100 W)
Nominal power
—
—
100%
Radiation shield stage
Heater range
Off
Off
Off
On (50 W)
Nominal power
—
—
—
100%
TABLE 4-4
General CCR and electronic control settings
Sample stage
Base to 20 K
20 K to 100 K
100 K to 250 K
250 K to 350 K
Nominal power
10% at 10 K
13% at 50 K
21% at 200 K
33% at 350 K
Heater range
Med
High
High
High
Proportional (P)
100
100
100
100
Integral (I)
25
25
25
25
Derivative (D)
0
0
0
0
Heater power values are based on a 4-arm configuration with probe mounts tied to the sample stage.
TABLE 4-5
Base temperature and up electronic control settings for the sample stage