65
|
www.lakeshore.com
3.6.4.3 Connecting the Vision System
1. Connect the microscope electronics. These consist of two parts:
a. The 12 V DC power supply connects to the plug on top of the microscope.
b. S-video cable: connect one end to the top of the microscope as well. It will
align in only one direction for the connection (FIGURE 3-20). Attach the other
end of this cable into the s-video connection on the monitor.
2. Attach the monitor power supply to the monitor. This is a 12 V DC connection.
3. Connect the optical fiber cable. The steps vary depending on whether you are
using the ring light or the coaxial light.
Ring Light:
a. Screw the light source adapter onto the bottom of the microscope, using the
three screws supplied (FIGURE 3-21). Finger tighten only.
b. Insert the fiber optic cable into the light source. Finger tighten only
c. Connect the power cord to the light source (see TABLE 3-1 for power
requirements).
FIGURE 3-20
Attach the s-video cable to the microscope
9
9
209
999
2022
3
3.5
3.5
.75X
Hand dial
Zoom
Quick release handle
Shaft collar
Focus