5.3.6 Reconfiguring
Microwave Cables
If you will be changing cables frequently, it is recommended to have a probe arm
available with the appropriate cable already installed, then switch probe arms using
the instructions in section 5.3.3 and section 5.3.4. However, if a cable change is nec-
essary, you can follow this procedure to do so.
5.3.6.1 Removing a Microwave Cable
Semirigid cables are used with microwave probes. Follow this procedure to remove a
semirigid cable from a probe arm assembly.
1. Remove the micro-manipulated stage using the instructions in section 5.3.2.
2. Remove the probe arm and base using the instructions in section 5.3.3.
3. Cut the unwaxed dental floss that ties the cable to the probe arm thermal anchor.
4. Using a 2.5 mm hex driver, remove the bottom set of four M3 screws that attach
the cable feedthrough assembly to the probe arm base (FIGURE 5-19).
5. Pull the cable feedthrough assembly and its attached cable out of the probe arm.
The semirigid cable has bends that require some reorientation of the cable feed-
through assembly as the cable is removed.
5.3.6.2 Installing a Microwave Cable
Semirigid cables are used with microwave probes. Follow this procedure to install a
semirigid cable onto a probe arm assembly.
1. Clean the o-ring groove of the cable feedthrough in the probe arm base. Clean,
inspect and lightly grease the o-ring with vacuum grease and place it in the
groove.
2. Carefully insert the cable into the probe arm base. Continue inserting the cable,
reorienting the cable feedthrough assembly, until it seats against the
probe arm base.
3. Using a 2.5 mm hex driver, attach the cable feedthrough assembly to the probe
arm base with the bottom set of four M3 screws (FIGURE 5-19).
4. Test fit a microwave probe onto the probe arm (section 3.7.4). Do not force the
threading of the plug if it does not tighten smoothly.
5. If the cable length seems inappropriate, use the instructions in section 5.3.6.3 to
adjust the cable length.
6. Remove the microwave probe you used for a test fit.
7. Tie the cable to the probe arm thermal anchor using unwaxed dental floss
8. Install the probe arm and base using section 5.3.4.
9. Install the micro-manipulated stage on the probe station using the steps in
10. Install a microwave probe using section 3.7.4.
FIGURE 5-19
Remove the bottom set of four M3 screws