147
|
www.lakeshore.com
6.4 Service
Reference
This is the service reference section.
6.4.1 Power
Requirements
Electrical power is required for the operation of the Model 336 temperature
controller, vision system, turbo pumping system, and CCR compressor. The CRX-6.5K
requires 1-phase power only. Most equipment is designed to operate over a range of
line voltages. Some equipment must be configured to operate at a specific voltage
within the range listed. This equipment is configured at Lake Shore to the voltage
specified when the equipment is ordered.
The electrical equipment can be grouped as shown in the tables below to distribute
the system power requirements over multiple facility circuits. TABLE 6-2 details the
operational power requirements for each circuit, and TABLE 6-3 details the
operational power requirements for the CCR compressor. The customer must provide
wiring that follows local safety codes to the CCR compressor.
The turbo pumping station should be powered from a separate circuit to isolate
electrical noise from the sensitive electronic instruments. The circuit protection used
for the turbo pumping system should be capable of handling high inrush currents.
The CCR compressor will draw high inrush currents when powered on. Circuit
protection capable of handling these short duration inrush currents is required to
prevent nuisance tripping of the safety devices. Thermal circuit breakers perform best
when available. The necessary utility connections are described in section 3.3.1.
The electrical requirements tables below summarize the individual circuit
requirements needed for the CRX-6.5K probe station.
Probe station equipment
Operational current required
100 VAC
120 VAC
220 VAC
240 VAC
Circuit 1
Model 336 temperature controller
2.5
2.1
1.2
1.1
Vision system
(monitor, camera and light source)
7.2
6.0
3.3
3.0
Circuit 2
TPS-FRG turbo pumping station
2.6
2.2
1.2
1.1
TABLE 6-2
1
-phase power requirements
Probe station
equipment
Power line voltage (±10%)
Operating current
Starting current
Recommended
circuit breaker
Power requirement
Circuit 3
HC4E1
1 phase,
208/230 V~ (± 5%), 60 Hz
200 V~ (± 5%), 50 Hz
230/240 V~ (± 5%), 50 Hz
Max 13/16 A
(50/60 Hz)
68 A
(50/60 Hz)
20 A
3.0 kW at 208 V, 60 Hz
2.6 kW at 200 V, 50 Hz
TABLE 6-3
1-phase power requirement for CCR