24
CHAPTER 2:
System Overview
Model CRX-6.5K Probe Station
2.3.8 High Temperature
Option (PS-HTSTAGE)
The Model CRX-6.5K probe station can be operated above 350 K using the high tem-
perature (HT) sample stage option. The HT sample stage is a coaxial stage that is
designed for 51 mm (2 in) diameter wafers and is easily removed from the
Model CRX-6.5K probe station similar to one of the other types of sample holders.
Electrical connection to the coaxial stage may be made through the standard
Model CRX-6.5K FT-BNC or FT-TRIAX option. The HT stage has two 50 W heaters
(100 W total) that are arranged in parallel; the combined parallel heater resistance is
25
)
. The HT stage has a platinum sensor embedded in the top sample plate for oper-
ation between 20 K and 675 K in the Model CRX-6.5K. Refer to section 5.2.2 for high
temperature stage operation.
Prior to June of 2013, the Model CRX-6.5K refrigerator needed to be configured at the
factory with the electrical wiring for operating the HT sample stage. After that date, the
Model CRX-6.5K refrigerator is pre-wired for upgrade to the high temperature stage.
2.3.9 Pump Line
Vibration Isolator
(PS-PLVI-40)
Vacuum pumps are a common source of vibration that can impact sensitive
measurements. A pump line vibration isolator can minimize the vibration from the
turbo vacuum system.
The PS-PLVI-40 pump line vibration isolator is recommended for isolating the turbo
pump for this application. It includes a bucket with NW 40 fittings, a 1 m flexible
stainless steel vacuum line and clamps.
2.3.10 Recirculating
Chiller Option
(RC-EM10)
Consult Lake Shore for the most current chiller information.
PS-PKIT
Included
accessories
Vacuum
line
Size
NW 40
Length
1 m (for specified system performance)
Type
Flexible stainless steel
Clamps
Size
NW 40, NW 25, NW 16
Reducers
Size
NW 40 to NW 25, NW 25 to NW 16
T
Size
NW 40
TABLE 2-2
Vacuum option components