Keysight N9927-90001 User’s Guide 281
Interference Analyzer (SA Mode) – Option 236
Record/Playback (SA Option)
—
FreqMaskTrigger OFF ON
Sets the power level which will trigger a recording. Set these power levels
(mask) exactly as you would set Limit Lines. (See
.) With
Record ON
, when any data point measurement is higher
than an upper limit, or lower than a lower limit, recording begins. See
Frequency Mask Trigger Configuration in the following section to learn how
to configure the mask.
Typically you would use FreqMaskTrigger with Segment Count, but it can also
be used with Record Time and Record Interval.
When used with
Segment Count
:
— When a mask limit is exceeded, subsequent traces are recorded until the
specified segment count is reached. Once segment count is reached,
exceeding another mask limit will again cause traces to be recorded until
the segment count is reached, and so forth.
— To record only those traces that exceed the limit, set Segment Count to 1
— To record ALL traces after a limit is exceeded, set Segment Count OFF.
When used with
Record Time
, the timer counts down even when traces are not
being recorded (limits are not being exceeded). Recording starts when a limit is
exceeded and stops when the timer reaches 0.
When used with
Record Interval
, a trace is recorded when a limit is exceeded,
then no recordings during the specified ‘wait’ interval, then the next trace is
recorded, followed by another time delay, and so forth. This pattern repeats
indefinitely if Segment Count is OFF, or (if ON) until Segment Count is reached.
Frequency Mask Trigger Configuration (SA Mode Only)
Create or recall a Frequency Mask to be used with the FreqMaskTrigger
feature.
A mask that is in use during a recording is NOT saved with the recording and
does NOT automatically appear during playback. Like Limits Lines, it is saved
and recalled with a state file.
Although they are created in the same manner, Limit Line files and FreqMask
files can NOT be used interchangeably.
Frequency Mask Trigger only applies to SA mode.
The Mask file is saved to, and recalled from, the Save /Recall Storage
Device location – NOT from the Record Playback Storage Device location.
These are two different settings. Learn how to set the Save /Recall Storage
Device in
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......