Keysight N9927-90001 User’s Guide 179
SA (Spectrum Analyzer) Mode
SA Mode Settings
The individual alignments may be:
— Coupled (default) - when they are in the same state, or
— UnCoupled - when at least one of them is in a different state than the other
ones
— Executed on demand, all of them, using the
Align All Now
softkey.
Each of the individual alignments:
— could be in one of three states:
—
Off
- the alignment is not applied and is not updated automatically
—
Auto
(default) - the alignment will be updated over time and
temperature as needed, no user interaction required
—
Hold
- the unit holds the alignment values from the last
alignment-execution and does not execute another alignment
automatically
— can be updated on demand using the corresponding
Align Now
soft-key.
“Individual Alignments” on page 180
.
— The
All Alignment
soft-key has the following functionality:
— when the individual alignments are Coupled, the state of all of them
can be changed simultaneously using the
All Alignment
soft-key
sub-menus
— when the individual alignments are UnCoupled, they will be coupled
when using the
All Alignment
soft-key sub-menus and their state
will be simultaneously changed accordingly
All Alignment and Align All Now
How to set All Alignment and Align All Now
— Press
Cal 5
IMPORTANT!
It is recommended that you use
Align All Now
, because it
includes the individual alignment InstAlign
Amptd Align Now
, as well as,
Burst Align Now
, and
ChanEQ Align Now
. But, if you would prefer to have
manual control over the individual alignments, then refer to the
SA
Individual Alignment Control
setting in
System Settings.
The notification
AMPTD?
displayed on screen means the InstAlign amplitude alignment or
channel equalization is stale (if its state is Hold) and requires an update or it is in the OFF state.
This situation does not occur when the respective alignment is in the Auto state. To clear this
notification, press the
Align All Now
softkey.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......