Keysight N9927-90001 User’s Guide 295
Built-in Power Meter Mode – Option 310
Built-in Power Meter Settings
Built-in Power Meter Settings
Select Built-in Power Meter Mode before making any setting in this section.
How to select Built-in Power Meter Mode
— Press
Mode
— Then
More
— Then
Power Meter (Built-in)
Frequency/ Span / Step
Valid Built-in Power Meter measurements can be made over the entire
frequency range of the FieldFox. Although it is possible to enter a center
frequency below the FieldFox minimum frequency, the LO feed-thru of the
internal SA hardware will invalidate these Built-in Power Meter measurements.
The measurement span (also known as Channel Width and Integration BW) is
the frequency range, half above and half below the center frequency, over
which power is detected and displayed. The maximum measurement span is
100 MHz.
How to enter center frequency, step size, and span
The default frequency is 50 MHz.
— Press
Freq/Dist
— Then
Center
— Then type a value using the numeric keypad. Then press a multiplier key.
Learn about multiplier abbreviations
“Multiplier Abbreviations” on page 41
The current frequency (f) is shown on the FieldFox screen.
— Optionally, use the
arrows or rotary knob to quickly change Center
frequency. The size of the frequency step can also be changed. Press
CF
Step
then type a step size value using the numeric keypad, then select a
multiplier key.
— Optionally press
Span
Type a value using the numeric keypad. Then press a
multiplier key. The default span is 20 MHz. The maximum span is 100 MHz.
The current span setting is shown as the dividend in the digital display. For
example:
-10 dBm / 1.00 MHz
Please observe the SA RF Input port damage level.
+27 dBm (320 mW) average
+33 dBm peak (2 W) <10
μ
s
+50 VDC
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......