Keysight N9927-90001 User’s Guide 189
SA (Spectrum Analyzer) Mode
SA Mode Settings
How to use Frequency Counter
— Press
Marker
.
— Press Normal, Delta, or Peak, or select an existing marker to use as a
frequency counter.
— Locate the marker somewhere on a CW signal. It does NOT have to be on
the peak.
— Then
More
— Then
Freq Count ON
— With Freq Count ON, you can scroll through the existing markers that are
located on a discrete CW signal, and read the frequency of the signal peak.
However, each frequency count update requires another sweep. If the
FieldFox is in HOLD, press
Single
to update the frequency count results.
Time Zero Fixed Marker
This feature was created to allow recall of vintage instrument states (older than Rev.
7.0) that included Zero span sweep with a trigger delay and at least one marker. Before
Rev. 7.0, these instrument states were saved and recalled with the equivalent of the
ON state of this setting. The default setting is OFF. Learn more about Triggering with
negative delay in
.
— Press
Mkr->/Tools
— Then
More
— Then
Mkr Properties
— Then
Time Zero Fixed
— Then choose from the following:
—
ON
Time Zero is fixed at the left-edge of the screen. A marker in
negative delay resides at this point. This is the vintage behavior.
—
OFF
(default) Time Zero is located at the T-zero indicators. A marker in
negative delay appears to the left of T-zero.
Audio Beep at Marker Power
When enabled, the Audio Beep feature emits a repetitive beep sound which
varies in tone pitch and repetition rate to indicate the relative power level of the
active marker. The highest tone pitch and fastest beep rate occurs when the
marker Y-axis position is at the top of the display. Conversely, the lowest pitch
and repetition rate occurs at the bottom of the display. Therefore, it is
important to scale the signals that you intend to measure between the top and
bottom of the screen. Learn more about Scale in
.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......