Point
- Each data point is measured <n> times and averaged before
going to the next data point. On subsequent sweeps, averaging
restarts by measuring each data point again <n> times. The average
counter is not updated because data is not displayed until all the
averages have been applied.
—
Point averaging is usually faster than sweep averaging.
However, you
may need to increase the Point Average count to obtain the same
level of noise reduction as with sweep averaging.
— While averaging is in process, press
Sweep 3
then
Restart
to restart the
averaging at 1.
IF Bandwidth
The FieldFox converts the received signal from its source to a lower
intermediate frequency (IF). The bandwidth of the IF bandpass filter is
adjustable. Reducing the IF receiver bandwidth reduces the effect of random
noise on a measurement. However, narrower IF bandwidths cause longer
sweep times.
How to set IF BW
—
Press BW 2.
— Then
IF BW
— Then choose from the following:
—
10 Hz
|
100 Hz
|
1 kHz
|
10 kHz
|
100 kHz
More
30 Hz
|
300 Hz
|
3 kHz
|
30 kHz
Smoothing
Trace smoothing averages a number of
adjacent
data points to smooth the
peak-to-peak noise values on a displayed trace. The number of adjacent data
points that are averaged is known as the smoothing aperture. Aperture is set
by specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Smoothing is used in Group Delay measurements, although it can be used with
any NA formatEXCEPT Polar or Smith Chart. Learn more about NA Mode
formats, including Group Delay, in
When enabled,
Smo
appears on the FieldFox screen.
How to set Smoothing
—
Press BW 2
— Then
Smoothing ON OFF
— Then
Sm. Aperture
and enter a value between 0 and 25 (percent) using the
numeric keypad.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......