Keysight N9927-90001 User’s Guide 61
DTF (Distance to Fault) Measurements
DTF Measurement Settings
—
More
then
DTF (Lin)
Faults are displayed on the Y-axis in linear
(unitless) format.
DTF Start and Stop Distance
In DTF measurements, you set the physical length of cable or other device to
be tested. The FieldFox calculates the frequency range of the measurement
from this distance. The longer the cable to be tested, the lower the frequencies
that are used. You can also set the frequencies manually using the Frequency
Mode [Bandpass] setting.
How to set Start and Stop Distance
— With a DTF measurement present, press
Freq/Dist
.
— Then choose from the following:
—
Start Distance
Enter a value using the numeric keypad, the
arrows, or the rotary knob, then
Enter
. By default, the Start Distance
is set to 0 Meters. This means that the measurement will display
faults starting at the point at which calibration standards are
connected.
—
Stop Distance
Enter a value between the start distance and 5 km (or
16,404 ft.) using the numeric keypad, the
arrows, or the rotary
knob, then
Enter
.
Frequency Mode
All DTF measurements are made with frequency settings and, using Inverse
Fourier Transform (IFT), the time and distance to faults are calculated.
How to set Frequency Mode
With a DTF measurement present,
— Press
Meas Setup 4
— Then
Frequency Mode
— Choose one of the following:
—
Lowpass
The frequency range of a DTF measurement is set
automatically based on the Start and Stop Distances. Use Lowpass
mode when the DUT is a cable ONLY.
The start and stop frequencies for the measurement are always annotated
on the screen below the start and stop distances.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......