170 Keysight N9927-90001 User’s Guide
SA (Spectrum Analyzer) Mode
SA Mode Settings
— Press
Sweep 3
— Then
Trigger Settings
— Then
Trig Slope
— Then choose from the following:
—
Po
Sweep is triggered by the rising (positive) edge of signal.
—
Neg
Sweep is triggered by the falling (negative) edge of signal.
Trigger Delay
After a valid trigger signal is received, the sweep begins after the specified
Trigger Delay time.
To see the rising edge of a repetitive signal which is triggered on that edge,
enter a negative trigger delay value (also known as pre-trigger). Adjust the
sweep time to observe the pre-trigger time if necessary.
— Press
Sweep 3
— Then
Trigger Settings
— Then
Trig Delay
— Enter a value using the numeric keypad, the
arrows, or the rotary
knob.
— Then select a multiplier key or press
Enter
. Learn more in
.
In Zero span, you can use Trigger Position as an easy way to set Trigger Delay
by positioning the trigger event on the FieldFox screen. See
.
Trigger Level
Used with Video and RF Burst triggering, a sweep is initiated when an
incoming signal crosses this level. The units depend on the Units setting. Learn
more in
RF Burst Trigger Level
uses an alignment process which is performed in the
background to set the detected signal level accuracy. Learn more in
Amplitude Alignment” on page 182
.
Video Trigger Level
is a zero span signal
level comparison. Therefore, the sweep will trigger close to the displayed level
in zero span measurements. In non-zero span measurements, processing can
cause broadband signal energy to display at lower power levels than the
originating time domain signal. Therefore, you may need to set the trigger level
higher than the displayed level.
— Press
Sweep 3
— Then
Trigger Settings
— Then
Trig Level
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......