Keysight N9927-90001 User’s Guide 351
ERTA (Extended Range Transmission Analysis) Mode - Option 209
Start Connection Wizard
— On the Source unit, connect the Port 1 RF Output connector to the
input of a power splitter.
— Connect one power splitter output to the Source unit Port 2 RF Input.
— Connect the other power splitter output to the DUT input.
— Connect the DUT output to the Port 2 RF Input of the Receiver unit.
Simplified method (not shown):
— Omit the power splitter. Connect the Source unit Port 1 RF Output
connector to the DUT input.
The measurement is a raw B receiver
.
Specs do NOT apply.
STEP 3: Select LAN address of ERTA Partner
The Receiver unit is the master partner, annotated with
RCVR Master
.
The Source unit is the follower partner, annotated with
SRC Follower
.
On the Receiver unit:
— Press
ERTA Partner (not specified)
. The IP addresses of potential FieldFox
units are displayed. Select the IP address of the Source unit. If 10.10.10.1
appears, then NO FieldFox units were found. The LAN setting “Obtain IP
Address using” MUST be set to DCHP. Learn more in
. Also be sure that Opt. 209 has been installed correctly on the
Source unit.
— Press
Continue
.
STEP 4: Test Partnership
— Press
Test Partnership
. If the Source unit is NOT already in ERTA mode, it
will be started, which may take a several seconds. When successful, LAN
OK and Trigger OK will appear on the screen.
— If either the LAN or Trigger test fails, then press
Clear Test
, then
Test
Partnership
to retry the test.
— Optional: Press
Send Text Message
, then enter a message to send to the
Receiver unit. A LAN connection is necessary. For quick communication,
custom Keywords can be setup using the FieldFox labeler. Learn how in
Any of the following will end the partnership:
— Press
EXIT Partnership
— Select Global or Mode Preset
— Change Modes
— Recalling a non-ERTA State file. Recalling an ERTA State file can
re-establish a partnership if the Source unit is still on the network at the
same location.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......