Keysight N9927-90001 User’s Guide 111
Time Domain – Option 010
Time Domain (Transform) Settings
Stimulus (Mode)
— There are three variations on how the Time Domain transform algorithm is
applied to the frequency domain measurement. Each method has a unique
application.
—
Lowpass Impulse
- Highest resolution. Most useful for seeing small
responses in devices that pass low frequencies, such as cables.
—
Lowpass Step
- Easiest to identify inductive and capacitive discontinuities
in devices that pass low frequencies such as cables.
In both Lowpass modes, frequencies down to DC and negative frequencies are
extrapolated. Therefore, when either Lowpass mode is selected, Set Freq.Low
Pass is automatically applied to adjust the start frequency. Learn more about
Set Frequency Lowpass below.
When the Start Frequency or resolution is changed AFTER selecting a Lowpass
mode, then Stimulus mode is set to Band Pass and Transform is disabled.
—
Bandpass Impulse
- Easiest method - can be used with any frequency
sweep. Most useful for measuring band limited devices such as filters and
DC blocked cables. This mode does NOT show capacitive and inductive
reactance. For the same frequency span and number of points, band pass
mode has twice the impulse width, which hides closely spaced responses
degrading the response resolution.
Set Frequency Lowpass
USE ONLY IN LOW PASS MODES. The Low Pass Start frequency is adjusted so
that the Low Pass Stop frequency is at harmonics of the Low Pass Start
frequency. Start frequency is computed by the following formula:
Low Pass Start Frequency = Stop Frequency / Number of points.
Start/Stop Time
The following settings adjust the
display resolution
, allowing you to zoom IN
or OUT on a response. When the start or stop time is updated, they may be
automatically adjusted to limit the display to one alias-free response on either
side of zero time.
How to set Start and Stop time
— Press
Measure Setup 4
— Then
Transform
— Then
Transform Start Stop
This setting is made automatically when either Lowpass mode is selected.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......