Keysight N9927-90001 User’s Guide 339
VVM (Vector Voltmeter) Mode - Option 308
VVM Mode Settings
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
Single
button is pressed. Use
Single
to conserve battery
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
— Press
Meas Setup 4
— Then choose one of the following:
—
Single
Automatically sets Continuous OFF and causes the FieldFox to
make ONE measurement, then hold for the next
Single
key press. The
Hold
annotation changes to an arrow --> while the measurement
occurs.
—
Continuous OFF
and
Hold
both put the FieldFox in
Hold
mode.
—
Continuous ON
Makes continuous measurements. This is the typical
setting when battery power is not critical.
— You can also use
Run / Hold +/-
to toggle between Single and Continuous.
VVM Calibration
CalReady is accurate for most zeroed VVM measurements at frequencies
below about 6 GHz. For the very highest accuracy, especially above 6 GHz,
then perform a QuickCal or Mechanical Cal. Learn more in
.
Press
Zero OFF
before calibrating (sets Zero to OFF).
Zeroing
Zeroing, also known as Normalization, or Data/Memory, simply “Zeroes” the
Magnitude and Phase display when
Zero ON
is pressed. All subsequent
measurements are relative to the measurement that was made when Zero was
pressed.
When an instrument state is saved with Zero ON, the zeroing data is saved with
the state file, similar to how a calibration is saved with a state file.
— When performing a 1-Port Cable Trimming operation, a reference cable is
connected and measured. Then Zero is pressed, and subsequent cables are
measured and trimmed to match electrical length of the reference cable.
QuickCal is only available on N991xA/2xA/3xA models.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......