Keysight N9927-90001 User’s Guide 405
Noise Figure Mode (Option 356 – CPU2 Only)
How to Define a Noise Figure Measurement
——
More
opens additional ENR Spot Value softkey
menu choices
——
Γ
Spec. Style [95th %ile]
Opens a menu of
softkeys to set the
Γ
specification style.
Then choose
Maximum
,
95th %ile
(default),
80th
%ile
,
Median
,
Mean
, or
Fixed
.
——
Γ
Distribution [Rayleigh]
Opens a menu of
softkeys to set the type of
Γ
distribution.
Then select
Rayleigh
(default),
Fixed
, or
Uniform
in Circle
.
—
ENR Table Edit/Save/Recall
Opens the ENR table menu to
edit, save, or recall noise source values (Not active with
ENR Mode set to Spot).
Then choose:
——
Edit ENR Table
Opens the currently selected
table of noise source values (generic default
noise source values are populated for quick
check). Enter your noise source’s values for more
accurate measurements.
——
New ENR Table
opens a new ENR table for
entering and saving additional noise source
values.
——
Save ENR Table
Opens a menu of softkeys to
enable you to enter a unique noise source
filename.
——
Recall ENR Table
Opens a menu of file recall
softkeys and displays a list of stored noise source
files.
——
Storage Device [Internal]
Opens a menu of
softkeys to set the storage location for your files.
Then choose
Internal
(default),
SD Card
, or
USB
.
——
Back
to return to previous menu.
IMPORTANT!
Save your noise source table with a unique filename to avoid
loss of table data. All noise figure files are saved with *.enr type suffixes.
The FieldFox displays
Replace existing file?
, because the displayed
default filename is the one last saved. Enter a unique filename to avoid
loss of table data (i.e., all noise figure files are saved with *.enr type
suffixes).
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......