Keysight N9927-90001 User’s Guide 355
ERTA (Extended Range Transmission Analysis) Mode - Option 209
Configure Measurement Settings
Learn more about Trace Display States in
“Trace Display States (SA Mode)” on
.
Normalize
Normalize can be used to quickly provide ‘zeroing’ for any of the ERTA (B, R, or
G) traces. On the sweep AFTER selecting Normalize, Trace 1 is copied to
Trace4. It is then put into View state and subtracted from that input on
subsequent sweeps. Any trace (1, 2, or 3) which is set to process that same
input (B, R, or G) will therefore be processing normalized measurement data.
This continues as long as the frequency span, number of points, and the trace
input (B, R, or G) is NOT changed. If any of these settings change, Normalize is
deactivated.
How to Normalize the B trace
— Press
Meas Setup 4
— Then
Remote Source
— Then
Normalize ON
Source Tracking Offset and Offset Reversal
Source offset tracking is used to offset the Source frequency from the Receiver
frequency by a fixed offset at every tune frequency. This feature is used when
the DUT contains a frequency converter.
How to set Tracking Offset and Offset Reversal
— On the Receiver unit, press
Meas Setup 4
,
— Then
Remote Source
— Then
Advanced
— Then choose one or both of the following:
—
Tracking Offset
Enter an offset value.
—
OfsTrkReversal
Determines whether Receiver Freq is added to or
subtracted from the Offset Freq.
——
OFF
SourceFreq = RecFreq. Use to measure
Upconverters.
——
ON
SourceFreq = Offset - RecFreq. Use to measure
Downconverters. The source sweeps in Reverse direction.
The Offset Freq. MUST be larger than the Receiver Stop Freq.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......