188 Keysight N9927-90001 User’s Guide
SA (Spectrum Analyzer) Mode
SA Mode Settings
If the Detection method and Averaging type are set to Auto when you enable a
Band/Interval Power marker, the Detection method will change to Average
(RMS) and Averaging type will change to Power average. Other Detection
methods or Averaging type settings will usually cause measurement
inaccuracy. Learn more about Detection method in
, and Averaging Type in
Summary:
— In Zero span, an Interval Power Marker yields the average power over the
specified time interval.
— In non-Zero span, a Band Power Marker yields the total power over the
specified frequency range.
How to measure Band/Interval Power with a marker
— Press
Marker
to create or select a Normal or Delta marker to use to
measure Band/Interval Power.
— Then press
More
— Then
Marker Function
— Then
Band/Int Power
— Then
Band (or Interval) Span
— Then choose from the following:
—
Auto
Measures power over 5% of the displayed span, or ½ graticules.
The Band Span is automatically updated to remain at 5% when you
change the frequency span.
—
Man
Enter a frequency or time value using the
arrows or the
rotary knob, then press
Enter
. Or enter a value using the numeric
keypad and select a frequency or time suffix. The Band Span remains
at the frequency or time value that you set as the span changes.
Frequency Counter at Marker
—
Available in SA Mode ONLY.
Use any existing marker to make
highly-accurate frequency counter readings. For highest accuracy, lock the
FieldFox to a stable external frequency reference. Learn how in
.
When Frequency Counter is ON, the FieldFox uses background sweeps to
zoom on the signal, measure the signal peak with 1 Hz resolution, and display
the frequency of the signal peak in the marker annotation area. The marker
does not move to the signal peak.
When Freq Counter is ON, measurement sweeps are considerably slower.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......