86 Keysight N9927-90001 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
S21 and S12 transmission measurements are used to measure the loss or gain
through a DUT over a specified frequency range. Both ends of the DUT must be
connected to the FieldFox. The FieldFox signal source is transmitted out one of
the test port connectors, through the DUT, and into the other test port
connector
How to measure S-parameters
1.
Press
Preset
then
Mode Preset (Factory)
2.
Press
Measure 1
then choose from the following:
—
S11
Reflection measurement at port 1.
—
S21
Forward 2-port transmission measurement.
—
S12
Reverse 2-port transmission measurement. May require an
option.
—
S22
Reflection measurement at port 2. May require an option.
OR
select a multi-trace configuration. Learn more on
3.
Press
Freq/Dist
then either
Start
and
Stop
or
Center
and
Span
to enter a
Frequency Range for the measurement.
4.
Press
BW 2
then
IFBW
to select the IF Bandwidth for the measurement.
Narrower bandwidths require more time to sweep, but lowers trace noise.
5.
Press
Sweep 3
then
Resolution
to select the number of data points for the
measurement. More data points require more time to sweep.
6.
Press
Cal 5
to calibrate the measurement. Learn more in
7.
All other settings can be made AFTER calibration without compromising
measurement accuracy.
Mixed-Mode S-Parameters
In NA Mode ONLY, mixed-mode S-parameters (also known as Balanced
measurements) are available with Option 212.
Because the FieldFox has only two test ports, only reflection measurements are
available. Connect the balanced input or output of your DUT to the FieldFox
ports 1 and 2
For highest accuracy, a Full 2-port calibration is required.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......