46 Keysight N9927-90001 User’s Guide
CAT (Cable and Antenna Test) Mode
CAT Mode Settings
—
Cable Loss (1-Port)
1-port reflection measurement that displays the loss of
a transmission line. Learn more on
—
Insertion Loss (2-Port)
2-port transmission measurement that accurately
displays the loss through a cable or other device in dB. Both ends of the
cable must be connected to the FieldFox. NO phase information is included
in this measurement. Learn more at
“2-Port Insertion Loss Measurements”
. This feature is available only with an option on some FieldFox
models. For detailed information, please view the FieldFox Configuration
Guide at:
http://literature.cdn.keysight.com/litweb/pdf/5990-9836EN.pdf
—
DTF (Lin)
Distance to Fault in Linear format.
—
TDR (Lin rho)
The Y-axis of the display is linear, real, unitless values. A trace
without reflections shows as 0 (zero). Maximum reflections from an open or
short show as 1.
—
TDR (ohm)
The Y-axis of the display is impedance (ohms).
—
TDR & DTF
TDR and DTF are displayed.
Quick Settings Table
Both CAT and NA Modes allow you to view and change most relevant settings
from a single location. All of these settings are discussed in this chapter and,
unless otherwise noted, ALL of these settings can also be made using the
standard softkey menus.
How to view and change Quick Settings
— Press
Meas Setup 4
.
— Then
Settings
.
— Press
Next Page
and
Previous Page
to view all settings. If these softkeys
are NOT available, then all available settings fit on one page.
— To change a setting:
— Use the
arrows to highlight a setting.
— Then press
Edit
. The current setting changes to
yellow
.
— Some settings require you to press a softkey to change the value.
Otherwise, use the numeric keypad,
arrows, or rotary knob to
change the value.
— When finished changing a value, press
Done Edit
.
— Press
Dock Window
to relocate the Settings table to a position relative to
the trace window. The Dock Window setting persists through a Preset.
Choose from the following
—
Full
(Default setting) Only the Settings table is shown on the screen.
The trace window is temporarily not shown
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......