244 Keysight N9927-90001 User’s Guide
IQ Analyzer (IQA) – Option 351 (CPU2 Only)
I/Q Analyzer (IQA) Mode Settings
—
Res BW
(Spectrum only) – Sets the resolution bandwidth of the FFT
analysis. Changing this value changes the FFT window size, analysis
length, and the sweep time (data acquisition capture length). This
softkey is only available when IQA is set to Spectrum (Freq Domain).
When the resolution bandwidth value is set to Auto (default) the
FieldFox automatically sets the optimum resolution bandwidth value.
(Default: 200 kHz, Minimum: 0.1 Hz, maximum: 3 MHz).
—
Analysis Length
Sets the FFT analysis length that is the length of the
complex I/Q data points (I/Q–pairs) vector that is used for spectrum
analysis, expressed in the number of samples. (Couplings: In Auto,
this setting is directly coupled to the Res BW and to Span. Minimum:
8, Maximum: 524,288. Where each FFT length "point" is equal to 1 I
and 1 Q point. So, 237 is equal to 237 I and 237 Q points).
—
FFT Window
Sets the FFT filtering window (spectrum analysis
window) shape for the I/Q data. Default: Flat Top
Then select one of the following:
There is a significant difference between resolution BW in FFT
measurements and the resolution BW of a traditional "sweeping"
spectrum analyzer (SA) measurement. The resolution BW in FFT
measurements does not directly control the IF BW like a traditional SA
swept measurement. Resolution bandwidth is directly related to the
analysis length (window) and or data acquisition capture length (sweep
time).
FFT Window
Description
Uniform
No windowing
Triangular
Barlett windowing
Hanning
This is a Hann window, sometimes called Hanning or "raised
cosine".
Hamming
This window has lower side-lobe peaks than the Hann window,
but has worse main-lobe flatness and poor side-lobe roll-off.
Blackman
This window is similar to the three-term Blackman-Harris
window, but it has simpler fixed-form coefficients and better
side-lobe roll-off.
Blackman Harris 3
This is the three-term Blackman-Harris window. This has the
minimum side-lobe level for a three-term cosine window.
Blackman Harris
This is the four-term Blackman-Harris window. This has the
minimum side-lobe level for a four-term cosine window with
60 dB/decade roll-off.
Gaussian Top
This is a five-term cosine window that resembles a Gaussian
window.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......