136 Keysight N9927-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
of the measurement WITHIN the frequency range of the waveguide Cal Kit.
Otherwise, an error message will appear during the ‘Calculating Steps’ portion
of the calibration.
Waveguide Cal Kits
Keysight sells two waveguide Cal Kit series: the premium 11644A series and
the economy N9911X series. Both are available online at
Effective Velocity Factor
Velocity factor is the speed at which an electromagnetic signal passes through
the transmission medium relative to the speed of light. This value is important
when distance is being calculated in DTF measurements (CAT mode) and Time
Domain (NA mode).
When the media is waveguide, the velocity factor changes with frequency.
FieldFox calculates this ‘effective’ velocity factor automatically. However, the
settings are different for CAT mode and NA mode.
CAT Mode - How to make Waveguide settings
These settings are necessary ONLY when making DTF measurements.
— Press
Measure 1
then
Distance to Fault (dB)
to select a
DTF
measurement.
— Press
Meas Setup 4
— Then
Settings
(Learn how to use the Quick Settings table in
and change Quick Settings” on page 46
.)
— Set
Media
= Waveguide
. Frequency Mode =
BandPass
is automatically
selected for you.
— Scroll down to
Waveguide Definitions
. Select the
Waveguide Standard
being used. If your waveguide standard is NOT listed:
— Select
User Waveguide
. Then press
Done
.
— Then
DTF Cable Specifications
, then
Edit/Save/Recall Cables,
then
Edit Cable
.
— Scroll to set
Waveguide Definition
. The default setting is
VF Corr
=
Auto.
— Set the Min, Max, and Cutoff Frequencies.
— Press
Done
, then press
Back
.
—
Cable Correction = Auto
is the default setting. The Effective Velocity Factor
is calculated automatically based on the frequencies of the waveguide
standard. To override this setting, set Cable Corr =
Man
.
NA Mode - How to make Waveguide settings
These settings are necessary ONLY when your measurement requires electrical
delay or port extensions, or if using Time Domain Transform.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......