124 Keysight N9927-90001 User’s Guide
Calibration for NA, CAT, and VVM Modes
— The connector choices that are offered depend on the frequency
range. For example, if you have selected a frequency range that
exceeds 18 GHz, the only connector options you will be presented
with in QuickCal are 'other 50 ohm' or 'other 75 ohm'. This is because
3.5 mm connectors, which go beyond 18 GHz, are not supported with
QuickCal.
— For each DUT port:
— Select the connector type of the DUT using the
arrows or rotary
knob.
— Press
Change Gender
to change the gender of the DUT connector.
— Press
Next
… to make the selection.
— If present, press
Finish
.
— Select/Change Cal Type. The FieldFox always RECOMMENDS a Cal Type
that will quickly and accurately calibrate all of the displayed S-parameters.
The other Cal Types that are listed will NOT necessarily calibrate all of the
displayed S-parameters. Learn more about Cal Types in
.
— Using the
arrows or rotary knob, select a Cal Type, then press
Select
and
Finish
.
— Press
Start Calibration
No Connection
— Disconnect the DUT from the FieldFox.
— If a jumper cable or adapter is required to connect the DUT to the FieldFox,
connect those components to the FieldFox test ports connectors. The
effects of those components will be measured and removed during the
calibration, and only the effects of the DUT will be displayed in the
measurement results. These should be high-quality components!
— Do NOT connect anything at the cal reference plane where the DUT
connects.
— Press
Measure
Optionally Connect Load(s)
— Connect a Load standard at the point where the DUT will be connected and
press
Measure
under the following conditions:
— When a jumper cable has too much loss. If a ‘Cal Failed’ error
appears and you skipped the Load measurement, then perform the
Cal again but include measuring the Load.
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......