92 Keysight N9927-90001 User’s Guide
NA (Network Analyzer) Mode
NA Mode Settings
Format
Select the display format in which to present measurement results. This setting
can be changed at any time without affecting calibration accuracy.
The marker on screen readout can be changed to formats other than the
display format. Learn more in
How to select a NA Format
— Press
Measure 1
.
— Then
Format [current setting]
.
— Then choose from the following:
—
Log Magnitude
Displays magnitude in dB
—
Linear
Displays positive values only. Y-axis: Unitless (U) for ratioed
measurements; Watts (W) for unratioed measurements.
—
VSWR
Used mainly for S11 and S22. Displays unitless reflection data.
—
Phase
Displays phase in degrees. The trace ‘wraps’ every 360 degrees,
from +180 to –180, for easy scaling.
—
Smith
Used mainly for S11 and S22, which are plotted on a Smith
Chart. Displays series resistance and reactance.
—
More
then…
—
Polar
Used mainly for S11 and S22. Displays magnitude and phase
of the reflection coefficient.
—
Group Delay
Used mainly for S21 and S12. Displays signal
transmission (propagation) time through a device in seconds. The
Group Delay aperture is the current Smoothing aperture. Default is
1.5% of the X-axis.
——Learn how to set Smoothing aperture on
——Learn more about Group Delay measurements at the FieldFox
Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldF
—
Real
Displays only the real (resistive) portion of the measured complex
data. The Y-axis is Unitless. Often used for Time Domain
measurements.
Learn more about Display Formats in the FieldFox Supplemental Online
Help:
http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htm
Summary of Contents for N9913A
Page 6: ...6 ...
Page 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Page 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Page 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Page 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Page 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Page 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Page 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Page 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Page 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Page 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Page 565: ......