Setup mode
OPERATION MANUAL
Chapter 4
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
74
PROCEDURE
Connect a suitable drain hose to the
>Drain<
[8].
Connect a suitable drain hose to the
>Expansion vessel drain<
[9] (if present).
Stick both ends of the hoses into suitable containers (e.g. original canister, which is compatible
with the thermal fluid).
Not all temperature control units are equipped with the same combination of drain valves. Skip
the point if your temperature control unit is not equipped with the drain valve indicated.
Open the
>Drain valve thermal fluid<
[3]
by turning it counterclockwise (turn 90° left as far as it
will go).
Open the
>Expansion vessel drain valve<
[5]
by turning it counterclockwise (turn 90° left as far
as it will go).
Wait until the thermal fluid has drained from the external application via the pump chamber and
the draining hose into the container.
Have another suitable container (e.g. a tub) ready to catch the thermal fluid from
>Emptying of
residues<
[10] (if present).
Remove the knurled screw to the
>Drain<
[10] (if present). The remaining thermal fluid will flow
from the temperature control unit into the container as soon as you have opened the knurled
screw.
Drain the external application. The description for the drainage procedure can be found in the
documents you received with the application.
Open the connection
>Circulation flow<
[1].
Open the connection
>Circulation return<
[2]. Leave the temperature control unit open for a
while to dry out (without screw caps and with the drain valve open).
Connect the connection
>Circulation flow<
[1] with your application.
Connect the connection
>Circulation return<
[2] with your application.
Re-fit the knurled screw to the
>Emptying of residues<
[10].
Close the
>Drain valve thermal fluid<
[3]
by turning it clockwise (turn 90° right as far as it will
go).
Close the
>Expansion vessel drain valve<
[5]
by turning it clockwise (turn 90° right as far as it wil
l
go).
Remove the drain hose from the
>Drain<
[8].
Remove the drain hose from the
>Expansion vessel drain<
[9] (if present).
Remove the containers used to collect the thermal fluid. Check if the thermal fluid can be reused.
When disposing of the thermal fluid, observe on page 17 observe Section
.
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...