Introduction
OPERATION MANUAL
Chapter 1
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
16
The temperature control unit is manufactured for industrial use. The temperature control unit main-
tains the temperature of certain applications, including glass or metal reactors or other expedient
items in laboratories and industry. Flow-through coolers and calibration baths must be used only in
combination with Huber temperature control units. Thermal fluids suitable for the overall system
are used. The chilling and heating capacity is provisioned at the pump connections or - where pre-
sent - in the tempering bath. The technical specification of the temperature control unit is given in
the data sheet (from page 99 in section
). The temperature control unit must be installed,
configured and operated according to the handling instructions in this operation manual. Failure to
comply with the operation manual is deemed improper use. The temperature control unit conforms
to state-of-the-art technology and the recognized safety regulations. Safety devices are built into
your temperature control unit.
1.2.4
Reasonably foreseeable misuse
The temperature control unit is operated outside of its specifications
MATERIAL DAMAGE CAUSED BY PREMATURE AGING / MATERIAL FATIGUE OF THE PLATE HEAT
EXCHANGER
Do not
permanently
operate the temperature control unit together with
very dynamic
and
high
temperature changes recurring in
short intervals
.
The
guarantee will expire
if the temperature control unit is exposed to this type of application.
Huber offers tried and tested technical solutions for these types of applications.
Without an Ex px cabinet, the temperature control unit is
NOT
protected against explosion and must
NOT
be installed or put into operation within an ATEX Zone. When operating the temperature control
unit in conjunction with an Ex px cabinet, the information in the annex (Section ATEX operation) must be
observed and followed. This annex is only provided for temperature control units delivered with an Ex
px cabinet. If this annex is missing, please immediately contact the Customer Support of Huber (the
telephone number is provided on page 98 in Section
).
The temperature control units Unistat T305, T305 HT, T305w HT do not have a plate heat exchanger.
Use with medical devices (e.g. in Vitro diagnostic procedure) or for direct foodstuff temperature
control is
NOT
permissible.
The temperature control unit must
NOT
be used for any purposes other than temperature control in
accordance with the operation manual.
The manufacturer accepts
NO
liability for damage caused by
technical modifications
to the temper-
ature control unit,
improper handling
or use of the temperature control unit if the operation manu-
al is
not observed
.
1.3
Responsible bodies and operators – Obligations and
requirements
1.3.1
Obligations of the responsible body
The operation manual is to be stored where it is easy to access in close proximity to the temperature
control unit. Only adequately qualified operators (e.g. chemists, CTA, physicists etc.) are permitted
to work with the temperature control unit. Operators must be trained before handling the tempera-
ture control unit. Check that the operators have read and understood the operation manual. Define
precise responsibilities of the operators. Personal protective equipment must be provided to the
operators.
▪
The responsible body must install a condensation water / thermal fluid drip tray below the tem-
perature control unit.
▪
The use of a drain tray may be prescribed by national law for the installation area of the tempera-
ture control unit (including accessories). The responsible body must check and apply the applica-
ble national regulations.
▪
Our temperature control unit complies with all applicable safety standards.
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...