Service/maintenance
OPERATION MANUAL
Chapter 7
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
90
PROCEDURE
Free-standing models:
Turn off the temperature control unit.
Disconnect the temperature control unit from the power supply.
Close the customer’s shut-off valves in the cooling water supply and return lines.
Remove the panel around the
>Cooling water inlet<
[13] and
>Cooling water outlet<
[14]. In-
formation: Located immediately behind the
>Cooling water inlet<
[13] is the dirt trap.
Place a collecting container below the
>Cooling water inlet<
[13].
Carefully detach the cover (hexagonal).
Remove and clean the metal strainer located below.
Clean the metal strainer under running water.
Re-insert the metal strainer after cleaning work.
Carefully fasten the cover (hexagonal).
Remove the collecting container from below the
>Cooling water inlet<
[13].
Re-mount the panel around the
>Cooling water inlet<
[13] and
>Cooling water outlet<
[14].
Open the customer’s shut-off valves in the cooling water supply and return lines.
Connect the temperature control unit to the power supply.
Turn on the temperature control unit.
We also provide service training. Please contact the Customer Support. The telephone number can
be found on page 98, section
.
7.3.5
Emptying the drip tray (if present)
Only valid for circulation pumps with mechanical seal.
Non-compliance with the safety data sheet for the thermal fluid to be used
INJURIES
Risk of injury to the eyes, skin, respiratory tract.
The safety data sheet for the thermal fluid to be used must be read prior to using it and its
content must be respected.
Observe the local regulations/work instructions.
Wear your personal protective equipment (e.g. temperature-resistant safety gloves, safety
goggles, safety footwear).
Danger of slipping because floor and work area are contaminated. Clean the work station and
follow the instructions for the disposal of thermal fluid and material on page 17 in section
»Proper disposal of resources and consumables«
.
This is not an actual leak but residues that are necessary for the lubrication of the mechanical seal.
The exiting quantity of thermal fluid is dependent on the thermal fluid itself and the working tem-
perature of the temperature control unit. The drops in the drip tray usually evaporate in case of
thermal fluids with a high vapor pressure. The drops usually do not evaporate in case of thermal
fluids with a low vapor pressure (e.g. silicone oils). Therefore, these residues must be drained from
time to time.
PROCEDURE
Hold a suitable container, e.g. beaker, below the
>Drain drip tray<
[7]. The drips can be cleanly
collected in the beaker and properly discarded. When cleaning filling accessories, observe page
17 section
»Proper disposal of resources and consumables«
.
Open the
>Drain drip tray<
[7].
Collect the drips.
Close the
>Drain drip tray<
[7].
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...