Introduction
OPERATION MANUAL
Chapter 1
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
20
1.5
Exemplary illustrations of the cooling variants
1.5.1
Consequence of inadequate energy dissipation
Room air/cooling water
Consequences of, for instance, contamination of the liquefier fins, inadequate clearance between
temperature control unit to wall/bath wall, room air/cooling water too warm, cooling water differ-
ential pressure too low, suction strainer contamination: The admitted energy can no longer be fully
discharged to the room air/cooling water
Thermal fluid circuit
Consequence of inadequate energy dissipation from the thermal fluid: The thermal fluid can only
dissipate the energy from your application to a limited extent.
Example: Air and
water cooling
Air cooling: Air inlet
Water cooling: Water
connection