Setup mode
OPERATION MANUAL
Chapter 4
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
64
4.2.11
Sub-category: “Display parameters”
In this function, the set parameters are displayed. Depending on the previous setting, these will be
the “Automatic control parameters” or the “Manual control parameters”.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Display parameters”.
Tap on “OK” after you have read/checked the parameters.
4.2.12
Setting the setpoint thresholds
The limits for the minimum and maximum setpoints serve for the safety of your system. They should
be set for the application field of the thermal fluid before starting the first temperature control and
when changing the thermal fluid. The maximum setpoint limit limits the setpoint setting for the bath
or flow temperature. The minimum setpoint limit protects against high viscosity or freezing at low
temperatures. Setpoint settings are only available in the temperature range between the minimum
and maximum setpoint limit.
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Protection Options”.
Tap on the category “Setpoint Limits”.
Tap on the sub-category “Minimum Setpoint”.
Enter the new value, using the numeric keypad that appears.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the “Min. setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Min. Setpoint” again.
Tap on the sub-category “Maximum Setpoint”.
Enter the new value, using the numeric keypad that appears.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the “Max. Setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Max. Setpoint” again.
Check the set values for the minimum and maximum setpoint at any system change, especially
when changing the thermal fluid.
Overview of the
temperature thresh-
olds
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...