Interfaces and software update
Chapter 6
OPERATION MANUAL
V2.4.0en/22.07.20//17.12
Unistat® T300 Series
79
▪
„2. setpoint selective”:
An open contact causes a thermoregulation to its original setpoint. A
closed contact causes a thermoregulation to the second setpoint.
▪
“Internal / Process”:
If the contact is open, the device-internal temperature sensor is used for
control. If the contact is closed, the additionally connected external temperature sensor is used
for control. Switching the contacts open/closed or closed/open switches between these two tem-
perature sensors.
▪
“Temperature control on/off”:
The temperature control process starts when switching from an
open to a closed contact. The temperature control process switches off when switching from a
closed to an open contact.
▪
“Release”:
The temperature control process switches off, if the temperature control process is
active and the contact is switched from closed to open. The temperature control process is
not
switched on
, if contact is further switched from open to closed!
▪
“Reset Messages”:
All messages in the Pilot ONE are reset (if possible) if the enable signal chang-
es from “open” to “closed”. Resetting is delayed in the event of a fault until the temperature con-
trol unit is in stand-by mode. Messages can be reset as often as required, but faults can be reset
only 3 times.
▪
“Program 1 Start/Stop”:
The temperature control program 1 is started if the enable signal chang-
es from “open” to “closed”. The temperature control program is stopped when changing from
“closed” to “open”.
▪
“Fill level alarm”:
When switching from closed to open contact, an alarm is triggered if the fill
level is too low. For this purpose, a level sensor must be connected and placed in the
> sight
glass<
[23] or in the external application.
▪
“Fill level warning”:
When switching from closed to open contact, a warning is generated if the fill
level is too low. For this purpose, a level sensor must be connected and placed in the
> sight
glass<
[23] or in the external application.
The interface is specified as a digital input. Do not apply voltage or current.
Pin
Signal
1.3 E2
2
E1
6.2.1.3
Connector POKO (floating contact) alarm
Signal contact for external monitoring.
Notice the functions provided by PoKo in the category “Interfaces”. The potential-free contact
(PoKo) signals the state of the temperature control unit via the contact position. A closed operating
contact means readiness for operation. If a fault or an error occurs, the operating contact is opened
(this applies to the make contact between pin 1 and pin 2).
The following settings are offered:
▪
“Off”:
The POKO displays the OK state whenever the temperature control unit is ready for opera-
tion. The unit is ready for operation about 30 seconds after switching on, once the internal con-
troller check has been completed. The OK status is terminated by switching off the unit or in case
a fault occurs.
▪
“Internal temperature relative":
Set the upper and lower limit of a temperature range around
the setpoint by using the POKO “min. value” and the POKO “max. value”.
The potential-free contact indicates the condition, that the setpoint differs from the actual value
by a smaller difference than was defined by the temperature range. Exceeding the set range only
results in the potential-free contact to switch (away from the OK state), but does not result in fur-
ther reactions of the temperature control unit. The contact returns to the OK status if the actual
value is within that range.
Pin assignment (front
view)
Pin assignment
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...