Interfaces and software update
Chapter 6
OPERATION MANUAL
V2.4.0en/22.07.20//17.12
Unistat® T300 Series
83
6.3.1
Jack LEVEL (Com.G@te® external only)
For level monitoring in the
>Sight glass<
[23].
This connection enables you to connect an external float switch (Order No. 6152), which is posi-
tioned in the
>Sight glass<
[23], for monitoring the level of your externally closed application. Acti-
vation via a potential-free contact.
The interface is specified as a digital input. Do not apply voltage or current.
Pin
Description
1
Level test (bypass via contact 2 → “absence”)
2
Level – (GND)
3
Level + (normally open contact)
6.3.2
Connector POKO (floating contact) alarm
Information on the interface can be found on page 79 in Section
»Connector POKO (floating con-
.
6.3.3
Jack AIF Reg-E-Prog
The analog interface has a programmable input channel and 3 output channels.
The analog interface of Com.G@te® is programmed in the category “Interfaces”.
Pin
Description
Signal
1
Current output, T external
0
a)
/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω resistor.
2
Current output, setpoint
0
a)
/4 - 20 mA or 0 - 10 V
3
GND for analog outputs
GND
4
Analog input (programmable)
0
a)
/4 - 20 mA or 0 - 10 V
Power input: 200 Ω working
resistance
Voltage input: 100 kΩ input
resistance
5
Current output, freely programmable
0
a)
/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω resistor.
6
GND for analog input
GND
a)
Consult with our Customer Support.
Pin assignment (front
view)
Pin assignment
Pin assignment (front
view)
Pin assignment
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...