Service/maintenance
OPERATION MANUAL
Chapter 7
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
86
7.2.2
Replacing the “Unistat® Control ONE”
PROCEDURE
Turn off the temperature control unit.
Disconnect the temperature control unit from the power supply.
Remove the “Pilot ONE” (see page 85, section
).
Undo the fastening screw on the “Unistat Control ONE”.
Also pull the “Unistat Control ONE” away upwards.
Carefully insert the replacement “Unistat Control ONE”.
Fasten the replacement “Unistat Control ONE” with the fastening screw.
Install the “Pilot ONE” (see page 85, section
).
Connect the temperature control unit to the power supply.
Turn on the temperature control unit.
7.3
Maintenance
Cleaning/maintenance while the temperature control unit is operating
MORTAL DANGER FROM ELECTRIC SHOCK
Stop an ongoing temperature control process.
Turn off the temperature control unit.
Also disconnect the temperature control unit from the power supply.
Performing maintenance work not described in these operation manual
MATERIAL DAMAGE ON THE TEMPERATURE CONTROL UNIT
Please contact Huber for maintenance work that is not described in these operation manual.
Maintenance work not described in these operation manual is reserved for qualified specialists
trained by Huber.
Safety-relevant components may only be replaced by equivalent ones. The specified safety
values for the respective component must be observed.
7.3.1
Function check and visual inspection
Cooling*
Description
Maintenance
interval
Comment
Person re-
sponsible
A/W
Visually inspect
hoses and hose
connections
Prior to switching
on the tempera-
ture control unit
Exchange leaking hoses and hose
connections prior to switching on
the temperature control unit.
Please see on page 88 the section
»Replacing temperature control
.
Operating
company
and/or opera-
tors
Replacing the “Unistat
Control ONE”
Monitoring interval
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...