Setup mode
Chapter 4
OPERATION MANUAL
V2.4.0en/22.07.20//17.12
Unistat® T300 Series
51
4
Setup mode
4.1
Setup mode
Moving the temperature control unit during operation
SERIOUS BURNS/FREEZING OF THE HOUSING PARTS/ESCAPING THERMAL FLUID
Do not move temperature control units that are in operation.
4.1.1
Turning on the temperature control unit
PROCEDURE
Switch on the temperature control unit using the
>Mains switch<
[36].
Temperature control unit with additional >Appliance switch<
[37]
:
Switch on the Pilot ONE using the
>Appliance switch<
[37].
First, a system test is conducted to test the full functionality of the temperature control unit. All
sensors, the all-pole disconnecting power disconnect relay for the power supply and the power
electronics of the main heater and the main heater itself are checked. A message appears on the
>Touchscreen<
[88] in case of an error or if a warning is issued. Contact Customer Support if in
doubt (the phone number can be found on page 98 in Section
).
The following data entry is required only for:
a) initial startup
b) reseting the temperature control unit to the factory settings
(see page 48, section
).
After switching on the temperature control unit, tap on the required system language.
Confirm your choice by tapping on “OK”.
Tap on the thermal fluid used.
Confirm your choice by tapping on “OK”.
Read the message and confirm by tapping on “OK”.
Enter the volume using the number keypad that appears.
Confirm your entry by tapping on “OK”.
If your temperature control unit is not integrated into a network, confirm the preset IP address
(0.0.0.0) by tapping on “OK”. This skips the network settings.
Enter the required IP address using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Enter the required Subnet Mask using the number keypad that appears.
Confirm your entry by tapping on “OK”.
Tap on the required remote control mode.
Confirm your choice by tapping on “OK”.
4.1.2
Turning off the temperature control unit
PROCEDURE
Warm the thermal fluid to room temperature.
Stop the thermoregulation.
Temperature control units with additional >Appliance switch<
[37]
:
Switch off the Pilot ONE using the
>Appliance switch<
[37].
Switch off the temperature control unit using the
>Mains switch<
[36].
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...