Interfaces and software update
Chapter 6
OPERATION MANUAL
V2.4.0en/22.07.20//17.12
Unistat® T300 Series
81
6.2.2
Interfaces at the “Unistat® Control ONE” top side
6.2.2.1
Service interface
This interface is exclusively used by Huber service engineers for service purposes. An adapter cable
makes this interface a RS232 serial port.
6.2.2.2
Female RS232 Serial (with adapter cable)
A PC, a SPS or a Process Control System (PCS) can be connected to this jack for remote control of the
controller electronics. Before plugging in the cable, check the settings in the “Interfaces” category
and adjust if necessary.
The interfaces used must meet the specifications of the generally accepted standards.
Pin
Signal
Description
2
RxD
Receive Data
3
TxD
Transmit Data
5
GND
Signal GND
6.2.2.3
Connection jack for Pt100 process controller sensor
A temperature sensor located in the connected application (Pt100, 4-wire technology, Lemosa con-
nector) is connected to the Pt100 connection jack. The external actual temperature is then recorded
and the operating temperature of the temperature control unit is permanently calculated and ad-
justed.
Depending on the operating temperature, isolation losses and exothermic heat, the operating
temperature (flow temperature) of the application can be significantly less than the setpoint of the
application. In this context, safety-critical thresholds for the temperature control fluid must be
strictly observed.
The control results contained in the data sheet can only be achieved with
shielded
sensor leads.
We recommend the external Pt100 process control sensor from the Huber accessories program.
Standard interfaces
at the “Unistat Con-
trol ONE” top side.
Pin assignment (front
view)
Pin assignment
Pin assignment (front
view)
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...