Function description
OPERATION MANUAL
Chapter 3
Unistat® T300 Series
V2.4.0en/22.07.20//17.12
48
Confirm the selection of the file by tapping on “OK”.
Select the setting group to be loaded from the list. A multiple selection is possible.
Confirm your choice by tapping on “OK”.
Read the message on the
>Touchscreen<
[88].
Switch off the temperature control unit. The settings have been loaded to the temperature
control unit.
Remove the USB flash drive from the temperature control unit.
3.10.4
Restore factory settings
Use this function to reset the temperature control unit to various basic states. This is useful if you
want to reverse various settings relatively quickly.
Restoring the factory settings is
only
possible if the temperature control unit is
not
carrying out a
task. If a task is active, do not switch off the temperature control unit until the application allows
this. Restoring the factory settings cannot be revoked. Depending on the type of resetting to facto-
ry settings performed you may have to enter parameters (process safety, thermal fluid used, cut-
out value etc.) again.
“X” = Value is reset, “–” = Value is not reset
(A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) Program Control-
ler; (E) = Menu; (F) = Com.G@te
Description
(A)
(B)
(C)
(D)
(E)
(F)
Acoustics
X
X
X
–
–
–
Display functions
Warnings; Messages; Inactive Menu Items; Resolution Temperature
X
X
X
–
–
–
Display modes
X
X
X
–
–
–
AutoStart response
X
X
X
–
–
–
Limitations
DeltaT limit; max. heating variable; max. cooling variable; max. permis-
sible current consumption (for 230 V temperature control units)
X
X
X
–
–
–
Com.G@te (only if Com.G@te is connected)
Analog Interface (configuration input)
Meaning of AIF input; response on cable break; measurement range
setting
X
X
X
–
–
X
Calibration
–
–
–
–
–
X
In case of analog error; current / voltage switching; filter constant
X
X
X
–
–
X
Configuration Output
AIF output variable; Output range setting
X
X
X
–
–
X
Calibration
–
–
–
–
–
X
RS232/RS485 (Hardware RS, Baud rate, device address); ECS Standby;
POKO Alarm
X
X
X
–
–
–
Properties (other)
Change thermal fluid (thermal fluid, liter indicator, bypass usage);
cooling bath / cooling (CC-E only)
X
X
X
–
–
–
Sensor calibration
–
–
–
–
–
–
Automated compressor
X
X
X
–
–
–
Enter program
X
–
–
X
–
–
Pump settings
Pump speed setpoint; Pump pressure setpoint; Control Mode
X
X
X
–
–
–
Fluid properties
Thermal fluid; liter indicator
X
X
X
–
–
–
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...