Setup mode
Chapter 4
OPERATION MANUAL
V2.4.0en/22.07.20//17.12
Unistat® T300 Series
57
4.2.6
Optimum control parameters for optimum temperature control
If the adjustment of the temperature does not correspond to the quality of the illustrations shown
above, you can adjust the control parameters. With Huber temperature control units, there are
various ways of finding the optimum control parameters. Depending on the facilities of the tempera-
ture control unit, you can choose the following processes:
▪
Use ex-factory parameters
(standard)
▪
Estimate Control Parameters
(only useful for bath thermostats with E-grade Basic and
internal control)
▪
Fast Identification
(from E-grade Exclusive)
▪
With Preliminary Test
(from E-grade Exclusive)
4.2.7
Sub-category: “Select auto/expert mode”.
Use of the “Expert mode” without a thorough knowledge of I&C technology.
MATERIAL DAMAGE TO THE APPLICATION
Only use this mode if you have a thorough knowledge of I&C technology.
Here you can select whether the control parameters are set in the “Automatic mode” or in the
“Expert mode”. To enter settings in “Expert mode” requires in-depth knowledge of process measur-
ing and control technology. Incorrect or insufficient settings can severely impair the function of the
temperature control unit.
In “Expert mode”, “Configuration auto” is deactivated and only “Configuration manual” is possible.
This is how to change mode:
PROCEDURE
Go to the “Categories Menu”.
Tap on the category “Temperature Control”.
Tap on the category “TAC/Manual”.
Tap on the sub-category “Select auto/expert”.
Choose between the dialog entries “Automatic mode” and “Expert mode”.
Confirm your choice by tapping on “OK”.
4.2.8
Sub-category: “Configuration auto”
You can only select this menu entry if “Automatic mode” is set.
4.2.8.1
Sub-category: “Find parameters”
4.2.8.1.1
Dialog entry: “Fast Identification”
Considering the small effort, the “Fast Identification” function of the controlled system provides you
with very quick and reliable adapted control parameters. These control parameters achieve a fast
and very accurate tuning performance. The more complex but also more precise identification “With
Preliminary Test” is only very rarely required.
Do not make any changes to the temperature control unit and to the application (temperature control
unit / external application) after your system has been started. Changes include e.g. filling / draining
process chamber, changing the mixer speed, change of position of Pt100 process control sensor etc.
PROCEDURE
Before setting the control parameters, make sure that the temperature control unit has reached
the set setpoint and has been controlling the temperature at this setpoint for a few minutes. Do
not stop the temperature control.
Summary of Contents for Pilot ONE Unistat T3*0
Page 1: ...Unistat T300 Series...
Page 2: ......
Page 3: ...OPERATION MANUAL Unistat T300 Series...
Page 4: ......
Page 7: ...OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 7 Layout of the Home screen...
Page 99: ...Annex Chapter 9 OPERATION MANUAL V2 4 0en 22 07 20 17 12 Unistat T300 Series 99 9 Annex...