9. Impulse Bandwidth Adjustments
and using the DATA knob to place the signal
peak near the top of the graticule.
24. Press MARKER
then MARKER
25. Using the DATA knob, adjust
the
marker down one side of
the
displayed signal to the 7.3
point; CRT MKR annotation
indicates 0.430 X.
26. Adjust
10
for MKR A indication of 5.00
while
maintaining the marker at 0.430 X using the DATA knob. Refer to
Figure 4-8 for
the
adjustment location.
27. Press MARKER la]. Adjust the marker to the 7.3
point on the
opposite side of the signal (CRT MKR
A
annotation indicates 1.00
28. The CRT MKR
A
annotation now indicates the impulse bandwidth
of the 10
bandwidth. The impulse bandwidth should
be
10.0
fl.O
29. Key in
3
and
If necessary, readjust by pressing
and using the DATA knob to place the signal peak near the top of
the
graticule.
30. Press MARKER
and MARKER
31. Using the DATA knob, adjust the marker down one side of the
displayed signal to the 7.3
point; CRT MKR A annotation
indicates 0.430 X.
32. Adjust
3
for MKR A indication of 1.5
while
maintaining the marker at 0.430 X using the DATA knob. Refer to
Figure 4-8 for
the
adjustment location.
33. Press MARKER la). Adjust the marker to the 7.3
point on the
opposite side of the signal (CRT MKR A annotation indicates 1.00
34. The CRT MKR La] annotation now indicates the impulse
bandwidth of the 3
bandwidth. The impulse bandwidth
should be 3.00
Option 462 4-29