7. IF Gain Uncertainty
13. Press
14. Set
[
REFERENCE LEVEL
]
to -1.9
15. Press MARKER
Set
to 100 Hz.
16. Set the frequency synthesizer for an output power level of -3.9
Set the amplitude increment for 2
steps.
17. Press MARKER
SEARCH
),
18. Set the analyzer
LEVEL
]
and the frequency synthesizer
amplitude according to Table 2-10. At each setting,
note
the MKR
A amplitude and record it in the table.
2-10. IF Gain Uncertainty, 2
Steps
(
REFERENCE LEVEL
)
Frequency
Deviation
Synthesizer (MARKER A
Amplitude
Amplitude
-1.9
-3.9
0 (ref)
-3.9
-5.9
-5.9
-7.9
-7.9
-9.9
-9.9
-11.9
L
19. Set
[
REFERENCE LEVEL
)
to 0
20. Set the frequency synthesizer for an output power level of -2.00
Set the amplitude increment for 0.1
steps.
21. Press MARKER
[
PEAK SEARCH
),
22. Set the analyzer
(
REFERENCE
and the frequency synthesizer
amplitude according to
2-11. At each setting, note the MKR
A amplitude and record it in the table.
2-24 Performance Tests