![HP HP 8566B Adjustment Manual Download Page 257](http://html.mh-extra.com/html/hp/hp-8566b/hp-8566b_adjustment-manual_161563257.webp)
13. Sweep, DAC, and Main Coil Driver Adjustments
s t o p p e d
1
2 . 0 0
3 . 5 0 0 v
10.00 : I dc
i I......_.._.._..._
- 1 . 0 0 0 0 0 s
0 . 0 0 0 0 0 s
1 . 0 0 0 0 0 s
2 0 0
1.375 v
Figure 3-59. Properly Adjusted DC Levels Between Sweep Ramps
13. Adjust
AUX adjustment to align dc level of 3 dead time
steps with upper dc level of each preceding sweep ramp. Refer
to Figure 3-59 and Figure 3-60 for typical display of proper and
improper adjustment.
s t o p p e d
1
2 . 0 0
3 . 5 0 0 v
10.00 : I dc
. . ../
2 0 0
1.375 v
Figure 3-60. Improperly Adjusted DC Levels Between Sweep Ramps
14. Disconnect the oscilloscope (and universal counter) from the
spectrum analyzer.
Adjustments 3-l 17