1. Low-Voltage Power Supply Adjustments
,
Figure 3-3.
IF-Display Section Adjustments (SN 3004A and Above)
5. Verify that the -15 V indicator
(yellow LED) is lit.
6. Connect the DVM to
The DVM indication should be
-15.000 f0.050 V dc. The -15 V supply is referenced to the + 15
V supply; therefore, if the -15 V supply is out of tolerance, a
circuit malfunction is indicated.
7. Verify that the + 120 V indicator
(yellow LED) is lit.
On IF-Display Sections with serial number prefix 3001A and below,
indicator
is a + 100 V indicator.
8. Connect the DVM to
The DVM indication should be
+ 120.0 f3.0 V dc. The + 120 V supply is referenced to the + 15
V supply; therefore, if the + 120 V supply is out of tolerance, a
circuit malfunction is indicated.
On IF-Display Sections with serial number prefix 3001A and below,
the DVM indication should be + 100.0 f2.0 V dc.
9. Verify that the
V indicator
(yellow LED) is lit.
10. Connect the DVM to
The DVM indication should be
f0.050 V dc. The
V supply is referenced to the + 15
V supply; therefore, if the
V supply is out of tolerance, a
circuit malfunction is indicated.
Adjustments