6. Video Processor Adjustments
6. Set step attenuator to 120
DVM indication should be 0.0000
f0.0005 V dc. If DVM indication is out of tolerance, adjust
ZERO
on
Log Amplifier-Detector. See Figure 3-34
for the adjustment location.
7. Set the step attenuator to 0
8. Key in
(
REFERENCE LEVEL
)
and adjust DATA knob, and then the
front panel AMPTD CAL control, for DVM indication as close to
+ 1.000 fO.OO1 Vdc as possible.
9. Connect DVM to
10. Adjust
OS for a DVM indication of 0.000
Vdc.
See Figure 3-36 for the location of the adjustment.
V i d e o P r o c e s s o r
R 3 6
R 3 2
F S Z E R O
Figure 3-36. Location of Video Processor Adjustments
11. Connect the DVM to
12. Set the step attenuator to 120
13. Adjust
ZERO for a DVM indication of 0.000 fO.OO1 Vdc.
14. Set the step attenuator to 0
15. Adjust
FS for DVM indication of
0.001 Vdc.
16. Repeat steps 12 through 15 until
no
further adjustments are
required.
17. Set step attenuator to 40
18. Key in
LOG
[
ENTER
(
REFERENCE LEVEL
)
-50
19. Connect DVM to
Record DVM indication. Indication
should be approximately
Vdc.
Vdc
20. Decrease reference level to -60
using the step key.
21. Adjust
LG OS for DVM indication of
fO.OO1
Vdc greater than the DVM indication recorded in step 19. See
Figure 3-36 for location of adjustment.
Adjustments