3-3. F&tory-Selected Components (continued)
Reference Adjustment Range of Values
Function of Component
Designator Procedure
1.78 K to 13.3 K
Log fidelity
8.25 K to 82.5 K
filter temperature compensation
1.0 K to 6.81 K
filter temperature compensation
1 K-OPEN
filter temperature compensation
8
1.0 to 8.2
Sets adjustment range of SYM.
17
8
180 to 270
Sets adjustment range of LC CTR.
8
1.0 to 8.2
Sets adjustment range of SYM.
8
1.0 to 8.2
Sets adjustment range of SYM.
8
180 to 270
Sets adjustment range of LC CTR.
8
180 to 270
Sets adjustment range of LC CTR.
8
180 to 270
0 to 9.09
Matches amplitude of LC to XTAL bandwidths.
3.16 K to 8.25 K
Adjusts LC filter bandwidth.
6.19 K to 12.1 K
Adjusts crystal filter bandwidth.
383 to 825
Matches amplitude of LC to XTAL bandwidths.
6.19 K to 12.1 K
Adjusts crystal filter bandwidth.
1 K to OPEN
Sets level of + 10 V TC supply.
1 K to OPEN
Sets level of + 10 V TC supply.
0 to 100
Adjusts bandwidth shape in 10
bandwidth.
3.1 6 K to 8.25 K
Adjusts LC filter bandwidth.
6.19 K to 12.1 K
Adjusts crystal filter bandwidth.
909 to 2.73 K
Adjusts positive feedback.
100 K to
Sets adjustment range of LC amplitudes.
10
11
10
10
10
O-16
Sets adjustment range of FREQ ZERO COARSE.
1.62 K to 2.61 K
Sets 18.4 MHz Local Oscillator power.
1.33 K to 3.48 K
Adjusts
step.
472 to 1.62 K
Adjust
step.
215 to OPEN
Adjusts
step.
42.2 to 75.0
Adjusts level of 3 MHz output.
12
1
56 to 82
Centers first pole.
7
56 to 82
Sets adjustment range of second pole P K.
7
56 to 82
Sets adjustment range of third pole P K.
7
56 to 82
Sets adjustment range of fourth pole P K.
7
56 to 82
Sets adjustment range of fifth pole P K.
7
1.5 to 12.0
Centers first pole.
10.0 K to 17.8 K
Adjusts crystal filter bandwidth.
10.0 K to 17.8 K
Adjusts crystal filter bandwidth.
10.0 K to 17.8 K
Adjusts crystal filter bandwidth.
10.0 K to 17.8 K
Adjusts crystal filter bandwidth.
3-14 Adjustments