2. High-Voltage Adjustment (SN 3004A and Above)
11. Remove the protective cover from the
High-Voltage
Regulator Assembly. A label should be visible on the
Assembly.
is mounted on the
non-component side of the High-Voltage Regulator Assembly as
shown in Figure 3-14.) Record the voltage listed on the label for
use in step 15. In cases where more than one voltage is listed on
this label, record the value which is closest to -2400 Vdc.
V dc
With power supplied to the instrument,
is at a voltage
level of approximately -2400 V dc. Be extremely careful.
12. Connect the high-voltage probe to
See Figure 3-14 for
the location of the test point.
L A B E L
Figure 3-14. Location
Label and
Point
13. Reconnect ac line cords to both instrument sections. Set the LINE
switch to ON.
14. Wait approximately 30 seconds for the dc regulator circuits to
stabilize.
15. Adjust
HV ADJ for a DVM indication equal to the
calibration factor (calculated in step times the voltage labeled
on the top of the
High-Voltage Assembly (noted in step
11). See Figure 3-13 for the location of the adjustment.
V dc
E
XAMPLE
:
If the calibration factor calculated in step 8 is 0.00099, and
is labeled for -2400 V, then adjust
HV ADJ
for a DVM indication of:
0.00099 x (-2400 V) = -2.376 V dc
16. With the front-panel INTENSITY control fully counter clockwise,
wait approximately 10 minutes to allow the high-voltage supply to
stabilize and the CRT to normalize. This
turn-on will extend
3-44 Adjustments