23. Track and Hold Adjustments
T R A C K A N D H O L D
( B e n e a t h C o v e r )
Figure 3-103. Location of Track and Hold Adjustments
10. Key in
TRACE A
11. Adjust
OFFS POS until MARKER A level indication as
indicated by CRT annotation flickers back and forth between .OO
and
12. Key in
TRACE A (VIEW)
13. Adjust
OFS NEG until MARKER A level indication as
indicated by CRT annotation flickers back and forth between .OO
and
14. Key in
TRACE A
15. Remove short from between
and
or remove
the SMB short from
Reconnect cable 7 (violet) to
16. Connect the DVM to
Connect
ground to the IF
section’s casting.
17. Press
(
REFERENCE LEVEL
]
and adjust DATA knob and front-panel
AMPTD CAL adjust for a DVM indication of
fO.OO1 V dc
at
18. Disconnect DVM from instrument.
19. Key in
TRACE A
MARKER
MARKER
SWEEP
20. Adjust
T/II GAIN for GAIN for MARKER
A
level
indication as indicated by CRT annotation of 100 fO.l
21.
22. Adjust
GPOS for MARKER A level indication as
indicated by CRT annotation of 100 fO.l
23. Key in
TRACE A
24. Adjust
GNEG for MARKER A level indication as
indicated by CRT annotation of 100
Adjustments