6
Bandwidth Adjustments
3 MHz bandwidth
30 Hz to 1 MHz bandwidths
+
-0%, 10 Hz bandwidth
30
and 100
bandwidth accuracy figures only applicable
Relative Humidity,
C .
The 6
bandwidth for each resolution bandwidth setting is
measured with the MARKER function to determine bandwidth
accuracy. The CAL OUTPUT is used for a stable signal source.
None required
1.
2.
3.
4.
5.
6.
7.
Press
Connect CAL OUTPUT to RF INPUT.
Key in spectrum analyzer settings as follows:
[
CENTER FREQUENCY
)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
MHZ
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MHz
. . . . . . . . . . . . . . . . . . . . . . 3 MHz
[
REFERENCE
. . . . . . . . . . . . . . -10
Press SCALE LIN pushbutton. Press
(resolution
bandwidth) for units in
Adjust
[
REFERENCE LEVEL
)
to position peak of signal trace at (or just
below) reference level (top) graticule line. Press SWEEP
Press MARKER
and place marker at peak of signal trace
with DATA knob. Press MARKER la] and position movable marker
6
down from the stationary marker on the positive-going edge
of the signal trace (the MARKER A amplitude readout should be
-6.00
f0.05
It may be necessary to press SWEEP
and adjust
(
CENTER FREQUENCY
_)
to center trace on screen.
Press MARKER La] and position movable marker 6
down
from the signal peak on the negative-going edge of the trace (the
MARKER Aamplitude readout should be .OO
f0.05
The 6
bandwidth is given by the MARKER A frequency readout. (See
Figure 4-l.) Record this value in
4-l.
4-2 Option 462