5. Resolution Bandwidth Switching Uncertainty
None required
1.
2 .
3 .
4 .
5 .
Press
(For instruments with Option 462, refer to Chapter 4.)
3 MHz Bandwidth Filter Adjustments
21.4 MHz Bandwidth Filter Adjustments
Down/Up Converter Adjustments
(uncorrected; referenced to 1 MHz bandwidth; 20 to
C)
I
Resolution Bandwidth
f 2 . 0
f 0 . 5
fl.O
10 Hz
30
Hz
100 Hz to 1 MHz
3
MHz
The CAL OUTPUT signal is applied to the input of the spectrum
analyzer. The deviation in peak amplitude of the signal trace is then
measured as each resolution bandwidth filter is switched in.
Connect CAL OUTPUT to RF INPUT.
Key in the following control settings:
[
CENTER
. . . . . . . . . . . . . . . . . . . . . . . . . 100
MHZ
SPAN
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
REFERENCE LEVEL
]
. . . . . . . . . . . . . . . . . . . . . . . .-8
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MHz
Press LOG
[
ENTER
and key in 1
Press MARKER
SEARCH], [al.
Key in settings according to
2-6. Press MARKER
[
PEAK SEARCH
]
at each setting, then read the amplitude deviation
from the MARKER A readout at the upper right of the display. (See
Figure 2-7.) The allowable deviation for each resolution bandwidth
setting is shown in the table.
Performance Tests